Mass spectrometry (MS) is an analytical technique that is used to measure the mass-to-charge ratio of ions. The results are presented as a mass spectrum, a plot of intensity as a function of the mass-to-charge ratio. These spectra are used to determine the elemental or isotopic signature of a sample, the masses of particles and of molecules, and to elucidate the chemical identity or structure of molecules and other chemical compounds. Mass spectrometry is used in many different fields and is applied to pure samples as well as complex mixtures.
In a typical MS procedure, a sample, which may be solid, liquid, or gaseous, is ionized, for example by bombarding it with a beam of electrons. This may cause some of the sample’s molecules to break up into positively charged fragments or simply become positively charged without fragmenting. These ions (fragments) are then separated according to their mass-to-charge ratio, for example by accelerating them and subjecting them to an electric or magnetic field: ions of the same mass-to-charge ratio will undergo the same amount of deflection.[1] The ions are detected by a mechanism capable of detecting charged particles, such as an electron multiplier. Results are displayed as spectra of the signal intensity of detected ions as a function of the mass-to-charge ratio. The atoms or molecules in the sample can be identified by correlating known masses (e.g. an entire molecule) to the identified masses or through a characteristic fragmentation pattern.
History

In 1886, Eugen Goldstein observed rays in gas discharges under low pressure that traveled away from the anode and through channels in a perforated cathode, opposite to the direction of negatively charged cathode rays (which travel from cathode to anode). Goldstein called these positively charged anode rays “Kanalstrahlen”; the standard translation of this term into English is “canal rays“. Wilhelm Wien found that strong electric or magnetic fields deflected the canal rays and, in 1899, constructed a device with perpendicular electric and magnetic fields that separated the positive rays according to their charge-to-mass ratio (Q/m). Wien found that the charge-to-mass ratio depended on the nature of the gas in the discharge tube. English scientist J. J. Thomson later improved on the work of Wien by reducing the pressure to create the mass spectrograph.

The word spectrograph had become part of the international scientific vocabulary by 1884.[2][3] Early spectrometry devices that measured the mass-to-charge ratio of ions were called mass spectrographs which consisted of instruments that recorded a spectrum of mass values on a photographic plate.[4][5] A mass spectroscope is similar to a mass spectrograph except that the beam of ions is directed onto a phosphor screen.[6] A mass spectroscope configuration was used in early instruments when it was desired that the effects of adjustments be quickly observed. Once the instrument was properly adjusted, a photographic plate was inserted and exposed. The term mass spectroscope continued to be used even though the direct illumination of a phosphor screen was replaced by indirect measurements with an oscilloscope.[7] The use of the term mass spectroscopy is now discouraged due to the possibility of confusion with light spectroscopy.[1][8] Mass spectrometry is often abbreviated as mass-spec or simply as MS.[1]
Modern techniques of mass spectrometry were devised by Arthur Jeffrey Dempster and F.W. Aston in 1918 and 1919 respectively.
Sector mass spectrometers known as calutrons were developed by Ernest O. Lawrence and used for separating the isotopes of uranium during the Manhattan Project.[9] Calutron mass spectrometers were used for uranium enrichment at the Oak Ridge, Tennessee Y-12 plant established during World War II.
In 1989, half of the Nobel Prize in Physics was awarded to Hans Dehmelt and Wolfgang Paul for the development of the ion trap technique in the 1950s and 1960s.
In 2002, the Nobel Prize in Chemistry was awarded to John Bennett Fenn for the development of electrospray ionization (ESI) and Koichi Tanaka for the development of soft laser desorption and their application to the ionization of biological macromolecules, especially proteins.[10]
Components

A mass spectrometer consists of 5 components: sample inlet, ion source (or ionizer), mass analyzer, detector, data system. Each component can be varied somewhat independently from the other components, and the full mass spectrometer itself can be included as a component in an analysis pipeline, allowing great flexibility.
The sample inlet prepares the sample to be analyzed in a suitable state.
The ion source takes samples from the sample inlet, converts a portion of the sample into a stream of ions, and outputs it to the mass analyzer. There is a wide variety of ionization techniques, depending on the phase (solid, liquid, gas) of the sample and the efficiency of various ionization mechanisms for the unknown species. An extraction system removes ions from the sample, which are then targeted into the mass analyzer.
The mass analyzer takes an input stream of ions and separates (“analyzes”) its constituent ions according to their m/z, the mass-to-charge ratio.
The detector measures the value of an indicator quantity and thus provides data for calculating the abundances of each ion present. Some detectors also give spatial information, e.g., a multichannel plate.
Illustrative example
The following describes the operation of a simple mass spectrometer that uses sector type. Other types are treated below.
Consider a sample of sodium chloride (table salt). In the ion source, the sample is vaporized (turned into gas) and ionized (transformed into electrically charged particles) into sodium (Na+) and chloride (Cl−) ions.
Sodium atoms and ions are monoisotopic, with a mass of about 23 daltons (symbol: Da or older symbol: u). Chloride atoms and ions come in two stable isotopes with masses of approximately 35 u (at a natural abundance of about 75 percent) and approximately 37 u (at a natural abundance of about 25 percent).
The analyzer part of the spectrometer contains electric and magnetic fields, which exert forces on ions traveling through these fields. The speed of a charged particle may be increased or decreased while passing through the electric field, and its direction may be altered by the magnetic field.
The magnitude of the deflection of the moving ion’s trajectory depends on its mass-to-charge ratio. Lighter ions are deflected by the magnetic force to a greater degree than heavier ions (based on Newton’s second law of motion, F = ma).
The streams of magnetically sorted ions pass from the analyzer to the detector, which records the relative abundance of each ion type. This information is used to determine the chemical element composition of the original sample (i.e. that both sodium and chlorine are present in the sample) and the isotopic composition of its constituents (the ratio of 35Cl to 37Cl).
Sample inlet
The sample inlet provides a stream of neutral molecules, so that the ion source can ionize them. There are several classes.[11]: Ch. 3.2
The classical design produces the stream of molecules in the form of a low-pressure gas. It is low-pressure, so that molecules within the stream does not frequently collide with each other. Ions must travel without frequent collisions, because the instrument separates ions by their motion in electric and magnetic fields. Collisions disturb that motion.
Despite being low-pressure, the sample stream is still at a much higher pressure than the rest of the system, which is held in a greater vacuum. Consequently, the sample stream can only enter the ion source chamber via a small pinhole, called a molecular leak.
If the sample itself is already a gas, then the sample inlet can simply be a gas inlet with pressure valve. If the sample is a volatile liquid or solid, then it can simply be placed in a chamber, vacuum-pumped to a low pressure. If the sample is not sufficiently volatile, the sample inlet can be fitted within an oven. Be careful not to decompose the sample.
If the sample is too nonvolatile to be turned into a low-pressure gas, then one can use a direct probe. The direct probe is a probe with a tip of thin wire loop, pin, or a capillary tube. The sample is placed in the tip. The probe is inserted through a vacuum lock into the ionization chamber, so that the tip is close to the ion source. The probe can be heated to increase vapor pressure further. Such a system is effective for samples with vapor pressures lower than mmHg at room temperature.
The sample can be prepared by an earlier analyzer. For example, it can simply be the gaseous output of a gas chromatography machine as in GC-MS, or the liquid output of a high-performance liquid chromatography machine, as in HPLC-MS.
The recently developed “atmospheric pressure ionization” techniques allow us to entirely dispense with the sample inlet.
Ion source

The ion source is the part of the mass spectrometer that transforms the sample from the sample inlet into a stream of ions. There are many techniques with various dimensions for tradeoffs between different choices of ion sources, in terms of monetary cost, accuracy, precision, compatibility with other stages of spectroscopy, ease of use, etc.
For example, electron ionization (EI) gives a high degree of fragmentation, yielding highly detailed mass spectra which when skilfully analysed can provide important information for structural elucidation/characterisation and facilitate identification of unknown compounds by comparison to mass spectral libraries obtained under identical operating conditions. However, EI is not suitable for coupling to HPLC, i.e. LC-MS, since at atmospheric pressure, the filaments used to generate electrons burn out rapidly. Thus EI is coupled predominantly with GC, i.e. GC-MS, where the entire system is under high vacuum.
Two techniques often used with liquid and solid biological samples include electrospray ionization (invented by John Fenn[12]) and matrix-assisted laser desorption/ionization (MALDI, initially developed as a similar technique Soft Laser Desorption) by K. Tanaka[13] for which a Nobel Prize was awarded and as MALDI by M. Karas and F. Hillenkamp[14]).
Hard ionization and soft ionization

Hard ionization techniques are those that impart high quantities of residual internal energy in the subject molecule, even after ionization. Part of the energy would be dissipated by rupturing chemical bonds within the ion, producing many fragmentary radical ions. The resultant ions tend to have m/z lower than the molecular ion (other than in the case of proton transfer and not including isotope peaks). The most common example of hard ionization is electron ionization (EI).
The radical ions produced after the first ionization are called second-generation product ions. The products are then directed towards the mass analyzer by a repeller electrode. The ionization process often follows predictable cleavage reactions that give rise to fragment ions which, following detection and signal processing, convey structural information about the analyte.
Conversely, soft ionization techniques are those that impart little residual energy onto the subject molecule and as such result in little fragmentation. Examples include fast atom bombardment (FAB), chemical ionization (CI), atmospheric-pressure chemical ionization (APCI), atmospheric-pressure photoionization (APPI), electrospray ionization (ESI), desorption electrospray ionization (DESI), and matrix-assisted laser desorption/ionization (MALDI).
Electron ionization
In electron ionization (EI), energetic electrons are produced by a hot cathode: a wire filament heated by running electric current through it, producing energetic electrons by thermionic emission. The electrons are then accelerated towards an anode. The voltage difference between the cathode and the anode determines the energy of the electron stream.[11]: Ch. 3.3
Most organic compounds have ionization energy 8-15 eV, but empirically, the efficiency of ionization is too low unless the electron beam has an energy of 50-70 eV per electron. The spectral features, including fragmentation patterns, depend on the energy setting. Conventionally for EI, the electron energy is standardized to be exactly 70 eV. Consequently, if one wishes to compare their EI-MS results against standard databases, or produce results that can be added to standard databases, one must standardize their EI electron energy to 70 eV.[11]: Ch. 3.3
EI accepts as input a stream of gas from the molecular leak. The gas stream crosses the electron stream perpendicularly. The collision ionizes the gas stream. The ion stream then is accerelated by a succession of electrodes. Typically an ion carrying 1 e would be accelerated to 1-10 keV.[11]: Ch. 3.3
EI has many benefits for routine mass spectrometry of small organic molecules. It is cheap and robust, with reproducible spectrograms. Databases for EI spectrograms are widely available and covers many such molecules.[11]: Ch. 3.3
EI is unsuited for large molecules, such as most biomolecules. It is a hard technique, which tends to break down large molecule ions during flight, since large molecules are particularly easy to fragment. Furthermore, large molecules are hard to volatilize without being degraded by heat.[11]: Ch. 3.3
Chemical ionization
In chemical ionization (CI), the analyte is ionized by a chemical reaction with an ionized reagent gas (itself ionized by some technique, such as by EI). The analyte gas and the reagent gas intersect, and react. The reaction then produces ionized analyte fragments by various mechanisms, including proton transfer, electron transfer, and adduct formation. The ion is then accelerated electrostatically, as in EI. CI ion sources are similar to EI sources, and most modern mass spectrometers can switch from EI to CI mode in minutes.[11]: Ch. 3.3
To ensure efficiency, the amount of reagent is much higher than the analyte. Consequently, a large amount of reagent ions would end up in the mass analyzer. This is usually handled by only measuring the part of the spectrogram with m/z above those of the main species in the reagent ion stream.[11]: Ch. 3.3
Common ionizing reagents for CI-MS include methane, ammonia, isobutane, and methanol. The proton affinity of the reagent gas and of the sample should be matched to ensure efficient ionization. If the proton affinity of the reagent gas is too high, ionization is inefficient. If the proton affinity of the reagent gas is too low, fragmentation is excessive.[11]: Ch. 3.3
For example:[11]: Ch. 3.3
- When the reagent is methane, the predominant reaction is proton transfer M + CH5+ → (MH)+ + CH4.
- When the reagent is ammonium, the predominant reaction is proton transfer M + NH4+ → (MH)+ + NH3.
- When the reagent is isobutane, the reaction may be proton transfer M + (CH3)3C+ → (MH)+ + (CH3)2C=CH2 or adduct formation M + (CH3)3C+ → (M · C(CH3)3)+.
- CI can also produce negatively charged analyte ions by proton abstraction or adduct formation. For example, ionized NF3 contains F−, which abstracts proton.
| Reagent gas | Proton affinity
(kcal/mole) |
Reagent ion(s) | Analyte ion(s) | Comments |
|---|---|---|---|---|
| H2 | 101 | H+3 | (MH)+ | Significant fragmentation |
| CH4 | 132 | CH+5, C2H+5 | (MH)+, (M · C2H5)+ | Some fragmentation, can form adducts |
| NH3 | 204 | NH+4 | (MH)+, (M · NH4)+ | Selective ionization, little fragmentation, can form adducts |
| (CH3)3CH | 196 | (CH3)3C+ | (MH)+, (M · C(CH3)3)+ | Selective protonation, little fragmentation. |
| CH3OH | 182 | CH3OH+2 | (MH)+ | |
| CH3CN | 188 | CH3CNH+ | (MH)+ |
Spray ionization

Electrospray ionization (ESI) and thermospray ionization (TSI) are suited for high molecular weight biomolecules and other labile or nonvolatile compounds, and especially in a LC-MS system. In both cases, the analyte is in a solution. The solution is sprayed out, by either electrospraying or thermospraying, into a stream of mist, which then evaporates into an ion stream. TSI has been supplanted by ESI for most purposes.[11]: Sec. 3.3
In TSI, the solution enters a heated capillary, producing a spray of droplets. The solvent evaporates, leaving ionized analytes. The ionization can occur due to 3 possible processes:
- The analyte is already in an ion form in the solution, and it is desorbed as the solvent evaporates.
- Acid-base transfer. If the solvent is a buffer solution, then the analyte can ionize by exchanging a proton with the buffer. This form of ionization is well-suited for LC-MS, since the liquid used in liquid chromatography already is a buffer solution.
- Plasmaspray ionization or filament-on operation. The solvent in the spray is electron-ionized by an energetic electron beam. The ionized solvent then chemically ionizes the analyte.
In ESI, the solution enters a capillary held at a high voltage (usually 2-5 kV), producing an electrospray into a heated chamber against a counterflow of a drying gas (usually nitrogen). This evaporates solvent molecules. The charge density increases on a droplet as it evaporates, until the electrostatic energy overcomes the surface tension energy, exploding the droplet. This process repeats, ending with an ion stream without the solvent. The ionization can occur by:[11]: Sec. 3.3
- Charge concentration in the droplets during evaporation.
- Electrochemical processes stemming from the electrostatic potential of the capillary.
Unlike in TSI, the analyte ions produced by ESI may carry a different amount of charge than the same analyte would carry in the solution. That is, an analyte that is naturally occurring as M+ in the solution may end up becoming a stream of ions containing a mixture of M+, M2+, M3+, etc. A large molecule can carry much charge. Typical proteins can carry many protons due to the presence of basic amino acid side chains, resulting in peaks at m/z = 600–2000 for proteins with a molecular weight ~200,000 Da. This is convenient, since it means the ion would not exceed the m/z limits on typical mass analyzers.[11]: Sec. 3.3
Desorption ionization
Desorption ionization (DI) is a family of ionization techniques. The main examples are secondary-ion mass spectrometry (SIMS), fast atom bombardment (FAB), and matrix-assisted laser desorption ionization (MALDI). In DI, the sample is dissolved or dispersed in a matrix, and bombarded by a high-energy (1-10 keV) beam of ions (SIMS), neutral atoms (FAB), or photons (MALDI). The beam ionizes and expels (desorbs) some sample molecules off the matrix. The ion is then accelerated electrostatically, as in EI.[11]: Sec. 3.3
The matrix should be nonvolatile and relatively inert, so that it would not react with the analyte. It should also be a good electrolyte to allow ion formation. Ionized fragments of the matrix itself forms usually have , so the spectrogram usually filters out that segment.[11]: Sec. 3.3
| Method | Beam | Common beam | Matrix | Typical ions | Analyte range |
|---|---|---|---|---|---|
| SIMS | ions | Ar+, Cs+ | nonvolatile electrolyte | (M + H)+, (M − H)−, (M + Na)+, (M + K)+ | <20 kDa |
| FAB | neutral atoms | Ar, Xe | nonvolatile electrolyte | (M + H)+, (M − H)−, (M + Na)+, (M + K)+ | <20 kDa |
| MALDI | photons | N2 laser 337 nm, infrared laser | UV-absorbing organic | M+ | <300 kDa |
SIMS and FAB are quite similar. SIMS uses an ion beam, usually Ar+ or Cs+. FAB uses a neutral atom beam, usually Ar or Xe. For SIMS and FAB, if the matrix compound is more acidic than the analyte, then predominantly (M + H)+ forms formed, and conversely (M − H)−. Also possibly forming (M + Na)+, (M + K)+, etc, if the matrix is contaminated incidentally (adventitiously) by sodium, potassium, etc. Typical matrix materials include glycerol, thioglycerol, 3-nitrobenzyl alcohol, diethanolamine, triethanolamine, and dithiothreitol–dithioerythritol mixture. They are usually used to analyze polypeptides and oligonucleotides up to 20 kDa.[11]: Sec. 3.3
MALDE uses a photon beam, usually the soft-UV 337 nm nitrogen laser. It can also use an infrared (IR) laser for direct analysis of samples contained in gels or thin-layer chromatography (TLC) plates. MALDE can analyze small polymers (~1 kDa), to oligosaccharides, oligonucleotides and polypeptides, antibodies, up to small proteins (~300 kDa). It is highly sensitive, requiring only femtomoles of sample.[11]: Sec. 3.3
Desorption/ionization on silicon (DIOS) is similar to MALDE, but without the matrix. The sample is deposited directly on a nanostructured (porous silicon) surface and the sample desorbed directly from the nanostructured surface through the adsorption of laser light energy. DIOS has been used to analyze organic molecules, metabolites, biomolecules and peptides, and, ultimately, to image tissues and cells.[15]
Inductively coupled plasma
Inductively coupled plasma (ICP) sources are used primarily for cation analysis of a wide array of sample types. In this source, a plasma that is electrically neutral overall, but that has had a substantial fraction of its atoms ionized by high temperature, is used to atomize introduced sample molecules and to further strip the outer electrons from those atoms. The plasma is usually generated from argon gas, since the first ionization energy of argon atoms is higher than the first of any other elements except He, F and Ne, but lower than the second ionization energy of all except the most electropositive metals. The heating is achieved by a radio-frequency current passed through a coil surrounding the plasma.
Photoionization mass spectrometry
Photoionization can be used in experiments which seek to use mass spectrometry as a means of resolving chemical kinetics mechanisms and isomeric product branching.[16] In such instances a high energy photon, either X-ray or uv, is used to dissociate stable gaseous molecules in a carrier gas of He or Ar. In instances where a synchrotron light source is utilized, a tuneable photon energy can be utilized to acquire a photoionization efficiency curve which can be used in conjunction with the charge ratio m/z to fingerprint molecular and ionic species. More recently atmospheric pressure photoionization (APPI) has been developed to ionize molecules mostly as effluents of LC-MS systems.
Ambient ionization
Some applications for ambient ionization include environmental applications as well as clinical applications. In these techniques, ions form in an ion source outside the mass spectrometer. Sampling becomes easy as the samples don’t need previous separation nor preparation. Some examples of ambient ionization techniques are Direct Analysis in Real Time (DART), DESI, SESI, LAESI, atmospheric pressure chemical ionization (APCI), desorption atmospheric-pressure chemical ionization (DAPCI), Soft Ionization by Chemical Reaction in Transfer (SICRT) and desorption atmospheric pressure photoionization DAPPI among others.
Other ionization techniques
Others include glow discharge, field desorption (FD), spark ionization and thermal ionization (TIMS).[17]
Mass analyzer
Mass analyzers separate the ions according to their mass-to-charge ratio.
There are many types of mass analyzers, using either static or dynamic fields, magnetic or electric fields. Each analyzer type has its strengths and weaknesses. Many mass spectrometers use two or more mass analyzers for tandem mass spectrometry (MS/MS). In addition to the more common mass analyzers listed below, there are others designed for special situations.
Theory
The following two laws govern the dynamics of charged particles in electric and magnetic fields in vacuum:
- (Newton’s second law of motion in the non-relativistic case, i.e. valid only at ion velocity much lower than the speed of light).
Here F is the force applied to the ion, m is the mass of the ion, a is the acceleration, Q is the ion charge, E is the electric field, and v × B is the vector cross product of the ion velocity and the magnetic field
Equating the above expressions for the force applied to the ion yields:
This differential equation is the classic equation of motion for charged particles. Together with the particle’s initial conditions, it completely determines the particle’s motion in space and time in terms of m/Q. Thus mass spectrometers could be thought of as “mass-to-charge spectrometers”. When presenting data, it is common to use the (officially) dimensionless m/z, where z is the number of elementary charges (e) on the ion (z=Q/e). This quantity, although it is informally called the mass-to-charge ratio, more accurately speaking represents the ratio of the mass number and the charge number, z.
Analyzer characteristics
There are several important analyzer characteristics.

The mass resolving power is the just detectable difference in m/z, at that particular m/z, usually measured measured in milli mass units (mDa). In particular, the resolution may be different at different m/z. It is usually operationalized as the full width at half maximum a peak at that given m/z.
The resolution is the ratio . It is unitless. A “high-resolution” mass analyzer typically has resolution >1 million. Concretely, that means it can resolve two peaks at ~1 kDa, differing by just ~1 mDa.
The absolute accuracy is the ratio of the m/z measurement error to the true m/z. It is defined as . The relative accuracy is . It is unitless.
The mass range is the range of m/z that a given analyzer can analyze.
The speed is the number of spectra lines per unit time that can be generated. Typically, higher resolution requires lower speed, and vice versa. Some mass spectrometers, such as the Fourier transform spectrometers, allow a continuous tradeoff between speed and resolution.
Sector

A sector-type mass analyzer uses a static electric and/or magnetic field to control the path of the ion beam. The most common type uses a magnetic field.
In a simplified case, the field is a constant field , and we have an ion of mass and charge that enters the field perpendicularly at velocity . Then, in the field, the ion undergoes circular motion of radius . If the ionizer has accelerated the ion under a known voltage , then the kinetic energy of the ion is , giving[11]: Sec. 3.4 Thus, the single input ion beam is analyzed into a fan of ion beams, the higher m/z ions turning at a larger radius. One can then select for one m/z by using a moving narrow slit. One can also use a fixed narrow slit, but vary or .
If there is just a single detector, then the analyzer can sweep across the mass spectrum one slit at a time. With multiple detectors, the speed of analysis increases.[18]
A pure magnetic sector analyzer can achieve resolution ~10,000. With an added electrostatic focusing field before or after the magnetic sector, the analyzer can achieve resolution ~100,000.[11]: Sec. 3.4
Particularly unstable ions may disintegrate while passing through the analyzer. If it undergoes the reaction Mz+ → M’z+ + (other parts) when entering the analyzer, it would undergo a circular motion of radius . This would then show up in the resulting spectrogram as a peak that is apparently made by an ion with . Since there is no actual ion with mass , this is called an abnormal peak, or a metastable ion peak. In general, a meta stable ion peak is a broad peak centered at a highly non-integer value of . The existence of a metastable ion peak at is strong evidence that two ions of masses are in fact related by the reaction Mz+ → M’z+ + (other parts).[11]: Sec. 3.5
Time-of-flight
The time-of-flight (TOF) analyzer takes as input a short pulse of ions, and lets it fly through a long vacuum tube. Ions with different m/z would arrive at the end of the tube at different times, thus analyzing them. The ions must enter the tube as pulses, because otherwise the TOF cannot be measured. If the ionizer outputs a continuous stream of ions, then it must be cut up into ion pulses, by methods such as a fast mechanical shutter, an oscillating electric field plus a narrow slit, etc. The MALDI ionizer is particularly well-suited for TOF analyzer, since the laser used in MALDI is usually a high-intensity pulse laser. MALDI-TOF has allowed measurement of mass spetrograms using only a few attomoles of material.[11]: Sec. 3.4
Concretely, let the length of the analyzer be , then the time taken for an ion of velocity to pass through the analyzer is . Since , we have[11]: Sec. 3.4 Often, the input ions are already concentrated in their m/z. That is, each ion in the input satisfies , where is a constant, and . In that case, we haveThis means that the signal can be measured by an oscilloscope in trigger mode. Indeed, the first GC-MS was done in 1950 using an oscilloscope. The experimenters took quick photos of the oscilloscope screen for each component exiting the gas chromatography machine.[19]
This distribution in velocities broadens the peaks shown on the count vs m/z plot, but will generally not change the central location of the peaks, since the starting velocity of ions is generally centered at zero. To fix this problem, time-lag focusing/delayed extraction has been coupled with TOF-MS.[20] If the ions do not start at identical kinetic energies, then some ions may lag behind higher kinetic energy ions decreasing resolution. Reflectron geometries are commonly employed to correct this problem.[21]
Longer tubes result in higher resolution. A reflector doubles the effective tube length. A reflector-TOF can reach a resolution of several thousands. For their low resolution, TOF mass spectrometers makes up for it in speed. They are particularly useful in chemical kinetics studies. It can study rapid reactions such as combustion and explosions.[11]: Sec. 3.4
Quadrupole mass filter

A quadrupole mass analyzers uses four parallel metal rods to create a quadrupole electrical field oscillating at radio frequency (RF), to selectively stabilize or destabilize the paths of ions passing through.
With each oscillation pattern, the ions passing through would undergo a helical motion. If the ion has m/z within a certain range, the helical motion is stable. Otherwise, the helical motion will increase in amplitude until the ion collides with the rods. This means that only ions with m/z within a narrow range can go through it. The full spectrogram can be swept through by changing the potentials on the rods, continuously or in a succession of discrete hops.
They are typically of low resolution ~1000, but they are cheap and small, so are good for “desktop” mass spectrometers.[11]: Sec. 3.4
The quadrupole mass filter is closely related to the quadrupole ion trap, particularly the linear quadrupole ion trap except that it is designed to pass the untrapped ions rather than collect the trapped ones, and is for that reason referred to as a transmission quadrupole. The quadrupole ion trap itself can be used as an analyzer, as in ion trap analyzers.
A common variation of the transmission quadrupole is the triple quadrupole mass spectrometer. The “triple quad” has three consecutive quadrupole stages, the first acting as a mass filter to transmit a particular incoming ion to the second quadrupole, a collision chamber, wherein that ion can be broken into fragments. The third quadrupole also acts as a mass filter, to transmit a particular fragment ion to the detector. It is a kind of MS/MS, and so it can perform various scan types characteristic of MS/MS.
A magnetically enhanced quadrupole mass analyzer has an additional magnetic field, axial or transverse.[22][23]
Ion traps
Three-dimensional quadrupole ion trap
The quadrupole ion trap works on the same physical principles as the quadrupole mass analyzer, but the ions are trapped and sequentially ejected. Ions are trapped in a mainly quadrupole RF field, in a space defined by a ring electrode (usually connected to the main RF potential) between two endcap electrodes (typically connected to DC or auxiliary AC potentials). The sample is ionized either internally (e.g. with an electron or laser beam), or externally, in which case the ions are often introduced through an aperture in an endcap electrode.
There are many mass/charge separation and isolation methods but the most commonly used is the mass instability mode in which the RF potential is ramped so that the orbit of ions with a mass a > b are stable while ions with mass b become unstable and are ejected on the z-axis onto a detector. There are also non-destructive analysis methods.
Ions may also be ejected by the resonance excitation method, whereby a supplemental oscillatory excitation voltage is applied to the endcap electrodes, and the trapping voltage amplitude and/or excitation voltage frequency is varied to bring ions into a resonance condition in order of their mass/charge ratio.[24][25]
Cylindrical ion trap
The cylindrical ion trap mass spectrometer (CIT) is a derivative of the quadrupole ion trap where the electrodes are formed from flat rings rather than hyperbolic shaped electrodes. The architecture lends itself well to miniaturization because as the size of a trap is reduced, the shape of the electric field near the center of the trap, the region where the ions are trapped, forms a shape similar to that of a hyperbolic trap.
Linear quadrupole ion trap
A linear quadrupole ion trap is similar to a quadrupole ion trap, but it traps ions in a two dimensional quadrupole field, instead of a three-dimensional quadrupole field as in a 3D quadrupole ion trap. Thermo Fisher’s LTQ (“linear trap quadrupole”) is an example of the linear ion trap.[26]
A toroidal ion trap can be visualized as a linear quadrupole curved around and connected at the ends or as a cross-section of a 3D ion trap rotated on edge to form the toroid, donut-shaped trap. The trap can store large volumes of ions by distributing them throughout the ring-like trap structure. This toroidal shaped trap is a configuration that allows the increased miniaturization of an ion trap mass analyzer. Additionally, all ions are stored in the same trapping field and ejected together simplifying detection that can be complicated with array configurations due to variations in detector alignment and machining of the arrays.[27]
As with the toroidal trap, linear traps and 3D quadrupole ion traps are the most commonly miniaturized mass analyzers due to their high sensitivity, tolerance for mTorr pressure, and capabilities for single analyzer tandem mass spectrometry (e.g. product ion scans).[28]
Fourier-transform ion cyclotron resonance

Fourier-transform mass spectrometry (FTMS), or more precisely Fourier-transform ion cyclotron resonance mass spectrometry (FT-ICR MS), measures mass by detecting the image current produced by ions spiralling in the presence of a magnetic field. Instead of measuring the deflection of ions with a detector such as an electron multiplier, the ions are injected into a Penning trap (a static electric/magnetic ion trap) where they effectively form part of a circuit. Detectors at fixed positions in space measure the electrical signal of ions which pass near them over time, producing a periodic signal. Since the frequency of an ion’s cycling is determined by its m/z ratio, this can be deconvoluted by Fourier transform on the signal.[29][30]
Ion cyclotron resonance (ICR) is an older mass analysis technique. It is similar to FTMS, in that the ions undergo cyclotron resonane. However, the ions are detected with a traditional detector. Ions trapped in a Penning trap are excited by an RF electric field until they impact the wall of the trap, where the detector is located. Ions of different mass are resolved according to impact time.
FTMS is generally the most sensitive analyzer on the market, achieving resolution ~1 million. It achieves such high resolution, since each ion is “counted” multiple times by the spectrometer. From the perspective of the uncertainty principle of signal processing, an oscillating signal that is highly delocalized in time is also highly localized in frequency.
Orbitrap

Orbitrap analyzers are similar to FTMS analyzers. Ions are electrostatically trapped in an orbit around a central, spindle shaped electrode. The electrode confines the ions so that they both orbit around the central electrode and oscillate back and forth along the central electrode’s long axis. This oscillation generates an image current in the detector plates which is recorded by the instrument. The frequencies of these image currents depend on the mass-to-charge ratios of the ions. Mass spectra are obtained by Fourier transformation of the recorded image currents.
Orbitraps have a high mass accuracy, high sensitivity and a good dynamic range. They are also quite small.[31]
High resolution
Some chemical species have the same chemical composition, but different isotopes (isotopomers). They can be distinguished by a very basic analyzer. Indeed, isotopes were discovered in 1922, very early in the history of mass spectrometry.
However, two chemical species can have the same number of nucleons (neutrons and protons). For example, C3 and SH4 both have 36 nucleons. They can still be distinguished, since proton and neutron mass differ by 1.4 mDa, and since different nuclei have different binding energy, which by results in a different mass. The net result is that C3 and S1H4 differ by 3.4 mDa. Even smaller differences are achievable by carefully matching isotopes. For example, C4 and S1H313C1 differ by 1.1 mDa. Such differences can be resolved by high resolution mass spectrometry, which reaches resolution >1 million, sufficient to resolve ~1 mDa difference at m/z ~ 1000. Note that the electron mass is 0.5 mDa.[32]
There are many techniques for high resolution MS,[33] but the highest resolution is achieved by FTMS with high magnetic field. With increasing magnetic field strength, the resolution and spectral acquisition speed increases linearly, while mass accuracy and dynamic range increases quadratically.[34] Consequently, the highest resolution is achieved by high field FTMS, up to 21 Tesla. It reaches resolution >2.7 million at m/z ~ 400, and mass measurement accuracy < 80 ppb.[35][32]
They are often employed in geochemical and petrochemical studies, since there is a lot of money in petroleum industry, and petroleum composition is highly complex.[32] Another application is in isotopic analysis of large proteins. Two proteins differing by a single isotopic atom are separated by , requiring a resolution of to resolve. For large proteins with m > 100 kDa and z ~ 50, such as monoclonal antibodies, this requires an analyzer that achieves resolving power > 100,000 at m/z ~ 2000.[35]
High resolution FTMS is also very sensitive as a detector. It has allowed detecting proteins in the 8–20 kDa range at a sample concentration of ~30 zeptomole (i.e. about 20,000 individual molecules).[36]
Detectors

The final element of the mass spectrometer is the detector. The detector records either the charge induced or the current produced when an ion passes by or hits a surface. In a scanning instrument, the signal produced in the detector during the course of the scan versus where the instrument is in the scan (at what m/Q) will produce a mass spectrum, a record of ions as a function of m/Q.
The linear dynamic range is the range over which the detector signal is linear with analyte concentration.
Typically, some type of electron multiplier is used, though other detectors including Faraday cups and ion-to-photon detectors are also used. Because the number of ions leaving the mass analyzer at a particular instant is typically quite small, considerable amplification is often necessary to get a signal. Microchannel plate detectors are commonly used in modern commercial instruments.[37] In FTMS and Orbitraps, the detector consists of a pair of metal surfaces within the mass analyzer/ion trap region which the ions only pass near as they oscillate. No direct current is produced, only a weak AC image current is produced in a circuit between the electrodes. Other inductive detectors have also been used.[38]
Tandem mass spectrometry

A tandem mass spectrometer is one capable of multiple rounds of mass spectrometry, usually separated by some form of molecule fragmentation. For example, one mass analyzer can isolate one peptide from many entering a mass spectrometer. A collision cell then stabilizes the peptide ions while they collide with a gas, causing them to fragment by collision-induced dissociation (CID). A further mass analyzer then sorts the fragments produced from the peptides. Tandem MS can also be done in a single mass analyzer over time, as in a quadrupole ion trap. There are various methods for fragmenting molecules for tandem MS, including collision-induced dissociation (CID), electron capture dissociation (ECD), electron transfer dissociation (ETD), infrared multiphoton dissociation (IRMPD), blackbody infrared radiative dissociation (BIRD), electron-detachment dissociation (EDD) and surface-induced dissociation (SID). An important application using tandem mass spectrometry is in protein identification.[39]
Tandem mass spectrometry enables a variety of experimental sequences. Many commercial mass spectrometers are designed to expedite the execution of such routine sequences as selected reaction monitoring, precursor ion scanning, product ion scanning, and neutral loss scanning.[40]
- In selected reaction monitoring, the first analyzer allows only a single mass through and the second analyzer monitors for multiple user-defined fragment ions over longer dwell-times than could be achieved in a full scan. This increases sensitivity.
- In product ion scans, the first mass analyzer is fixed to select a particular precursor ion (“parent”), while the second is scanned to find all the fragments (“products”, or “daughter ions”) to which it can be fragmented in the collision cell.
- In precursor ion scans, the second mass analyzer is fixed to select a particular fragment ion (“daughter”), while the first is scanned to find all possible precursor ions that could give rise to this fragment.
- In neutral loss scans, the two mass analyzers are scanned in parallel, but separated by the mass of a molecular subunit of interest to the analyst. Ions are detected if they lose that fixed mass during fragmentation. This can be used to look for any chemical that is capable of losing a particular neutral group, for example a sugar residue. Together, neutral loss and precursor ion scans can be used to hunt for chemicals with particular motifs.
Another type of tandem mass spectrometry used for radiocarbon dating is accelerator mass spectrometry, which uses very high voltages, usually in the mega-volt range, to accelerate negative ions into a type of tandem mass spectrometer.
The METLIN Metabolite and Chemical Entity Database[41][42][43][44][45] is the largest repository of experimental tandem mass spectrometry data acquired from standards. The tandem mass spectrometry data on over 960,000 molecular standards (as of October 2025)[41][44] is provided to facilitate the identification of chemical entities from tandem mass spectrometry experiments.[46] In addition to the identification of known molecules it is also useful for identifying unknowns using its similarity searching/analysis.[47] All tandem mass spectrometry data comes from the experimental analysis of standards at multiple collision energies and in both positive and negative ionization modes.[41]
Common mass spectrometer configurations and techniques
When a specific combination of source, analyzer, and detector becomes conventional in practice, a compound acronym may arise to designate it succinctly. One example is MALDI-TOF, which refers to a combination of a matrix-assisted laser desorption/ionization source with a time-of-flight mass analyzer. Other examples include inductively coupled plasma-mass spectrometry (ICP-MS), accelerator mass spectrometry, thermal ionization-mass spectrometry (TIMS) and spark source mass spectrometry.
Certain applications of mass spectrometry have developed monikers that although strictly speaking would seem to refer to a broad application, in practice have come instead to connote a specific or a limited number of instrument configurations. An example of this is isotope-ratio mass spectrometry, which refers in practice to the use of a limited number of sector based mass analyzers; this name is used to refer to both the application and the instrument used for the application.
Separation techniques combined with mass spectrometry
An important enhancement to the mass resolving and mass determining capabilities of mass spectrometry is using it in tandem with chromatographic and other separation techniques.
Gas chromatography

A common combination is gas chromatography-mass spectrometry (GC/MS or GC-MS). In this technique, a gas chromatograph is used to separate different compounds. This stream of separated compounds is fed online into the ion source, a metallic filament to which voltage is applied. This filament emits electrons which ionize the compounds. The ions can then further fragment, yielding predictable patterns. Intact ions and fragments pass into the mass spectrometer’s analyzer and are eventually detected.[48] However, the high temperatures (300 °C) used in the GC-MS injection port (and oven) can result in thermal degradation of injected molecules, thus resulting in the measurement of degradation products instead of the actual molecule(s) of interest.[49]
Liquid chromatography

Similarly to gas chromatography MS (GC-MS), liquid chromatography-mass spectrometry (LC/MS or LC-MS) separates compounds chromatographically before they are introduced to the ion source and mass spectrometer. It differs from GC-MS in that the mobile phase is liquid, usually a mixture of water and organic solvents, instead of gas. Most commonly, an electrospray ionization source is used in LC-MS. Other popular and commercially available LC-MS ion sources are atmospheric pressure chemical ionization and atmospheric pressure photoionization. There are also some newly developed ionization techniques like laser spray.
Capillary electrophoresis–mass spectrometry
Capillary electrophoresis–mass spectrometry (CE-MS) is a technique that combines the liquid separation process of capillary electrophoresis with mass spectrometry.[50] CE-MS is typically coupled to electrospray ionization.[51]
Ion mobility
Ion mobility spectrometry-mass spectrometry (IMS/MS or IMMS) is a technique where ions are first separated by drift time through some neutral gas under an applied electrical potential gradient before being introduced into a mass spectrometer.[52] Drift time is a measure of the collisional cross section relative to the charge of the ion. The duty cycle of IMS (the time over which the experiment takes place) is longer than most mass spectrometric techniques, such that the mass spectrometer can sample along the course of the IMS separation. This produces data about the IMS separation and the mass-to-charge ratio of the ions in a manner similar to LC-MS.[53]
The duty cycle of IMS is short relative to liquid chromatography or gas chromatography separations and can thus be coupled to such techniques, producing triple modalities such as LC/IMS/MS.[54]
Data and analysis

Data representations
Mass spectrometry produces various types of data. The most common data representation is the mass spectrum.
Certain types of mass spectrometry data are best represented as a mass chromatogram. Types of chromatograms include selected ion monitoring (SIM), total ion current (TIC), and selected reaction monitoring, among many others.
Other types of mass spectrometry data are well represented as a three-dimensional contour map. In this form, the mass-to-charge, m/z is on the x-axis, intensity the y-axis, and an additional experimental parameter, such as time, is recorded on the z-axis.
Data analysis
Mass spectrometry data analysis is specific to the type of experiment producing the data. General subdivisions of data are fundamental to understanding any data.
Many mass spectrometers work in either negative ion mode or positive ion mode. It is very important to know whether the observed ions are negatively or positively charged. This is often important in determining the neutral mass but it also indicates something about the nature of the molecules.
Different types of ion source result in different arrays of fragments produced from the original molecules. An electron ionization source produces many fragments and mostly single-charged (1-) radicals (odd number of electrons), whereas an electrospray source usually produces non-radical quasimolecular ions that are frequently multiply charged. Tandem mass spectrometry purposely produces fragment ions post-source and can drastically change the sort of data achieved by an experiment.
Knowledge of the origin of a sample can provide insight into the component molecules of the sample and their fragmentations. A sample from a synthesis/manufacturing process will probably contain impurities chemically related to the target component. A crudely prepared biological sample will probably contain a certain amount of salt, which may form adducts with the analyte molecules in certain analyses.
Results can also depend heavily on sample preparation and how it was run/introduced. An important example is the issue of which matrix is used for MALDI spotting, since much of the energetics of the desorption/ionization event is controlled by the matrix rather than the laser power. Sometimes samples are spiked with sodium or another ion-carrying species to produce adducts rather than a protonated species.
Mass spectrometry can measure molar mass, molecular structure, and sample purity. Each of these questions requires a different experimental procedure; therefore, adequate definition of the experimental goal is a prerequisite for collecting the proper data and successfully interpreting it.
Interpretation of mass spectra
Since the precise structure or peptide sequence of a molecule is deciphered through the set of fragment masses, the interpretation of mass spectra requires combined use of various techniques. Usually the first strategy for identifying an unknown compound is to compare its experimental mass spectrum against a library of mass spectra. If no matches result from the search, then manual interpretation[55] or software assisted interpretation of mass spectra must be performed. Computer simulation of ionization and fragmentation processes occurring in mass spectrometer is the primary tool for assigning structure or peptide sequence to a molecule. An a priori structural information is fragmented in silico and the resulting pattern is compared with observed spectrum. Such simulation is often supported by a fragmentation library[56] that contains published patterns of known decomposition reactions. Software taking advantage of this idea has been developed for both small molecules and proteins.
Analysis of mass spectra can also be spectra with accurate mass. A mass-to-charge ratio value (m/z) with only integer precision can represent an immense number of theoretically possible ion structures; however, more precise mass figures significantly reduce the number of candidate molecular formulas. A computer algorithm called formula generator calculates all molecular formulas that theoretically fit a given mass with specified tolerance.
A recent technique for structure elucidation in mass spectrometry, called precursor ion fingerprinting, identifies individual pieces of structural information by conducting a search of the tandem spectra of the molecule under investigation against a library of the product-ion spectra of structurally characterized precursor ions.[57]
Particularly unstable ions may disintegrate while passing through the analyzer. If it undergoes the reaction Mz+ → M’z+ + (other parts) when entering the analyzer, it would undergo a circular motion of radius . This would then show up in the resulting spectrogram as a peak that is apparently made by an ion with . Since there is no actual ion with mass , this is called an abnormal peak, or a metastable ion peak. In general, a meta stable ion peak is a broad peak centered at a highly non-integer value of . The existence of a metastable ion peak at is strong evidence that two ions of masses are in fact related by the reaction Mz+ → M’z+ + (other parts).[11]: Sec. 3.5
Applications

Mass spectrometry has both qualitative and quantitative uses. These include identifying unknown compounds, determining the isotopic composition of elements in a molecule, and determining the structure of a compound by observing its fragmentation. Other uses include quantifying the amount of a compound in a sample or studying the fundamentals of gas phase ion chemistry (the chemistry of ions and neutrals in a vacuum). MS is now commonly used in analytical laboratories that study physical, chemical, or biological properties of a great variety of compounds. Quantification can be relative (analyzed relative to a reference sample) or absolute (analyzed using a standard curve method).[58]
As an analytical technique it possesses distinct advantages such as: Increased sensitivity over most other analytical techniques because the analyzer, as a mass-charge filter, reduces background interference, Excellent specificity from characteristic fragmentation patterns to identify unknowns or confirm the presence of suspected compounds, Information about molecular weight, Information about the isotopic abundance of elements, Temporally resolved chemical data.
A few of the disadvantages of the method is that it often fails to distinguish between optical and geometrical isomers and the positions of substituent in o-, m- and p- positions in an aromatic ring. Also, its scope is limited in identifying hydrocarbons that produce similar fragmented ions.
Isotope ratio mass spectrometry

Mass spectrometry is also used to determine the isotopic composition of elements within a sample. Differences in mass among isotopes of an element are very small, and the less abundant isotopes of an element are typically very rare, so a very sensitive instrument is required. For the measurement of light elements (e.g. H,C,N,O,S), isotope ratio mass spectrometers usually use a single magnet to bend a beam of ionized particles towards a series of Faraday cups which convert particle impacts to electric current. A fast on-line analysis of deuterium content of water can be done using flowing afterglow mass spectrometry, FA-MS. Probably the most sensitive and accurate mass spectrometer for this purpose is the accelerator mass spectrometry. This is because it provides ultimate sensitivity, capable of measuring individual atoms and measuring nuclides with a dynamic range of ~1015 relative to the major stable isotope.[59] Isotopic signatures can serve as markers of a variety of processes. Some isotope ratios are used to determine the age of materials for example as in carbon dating. Labeling with stable isotopes is also used for protein quantification (see protein characterization below).
Membrane-introduction mass spectrometry: measuring gases in solution
Membrane-introduction mass spectrometry combines the isotope ratio mass spectrometry with a reaction chamber/cell separated by a gas-permeable membrane. This method allows the study of gases as they evolve in solution. This method has been extensively used for the study of the production of oxygen by Photosystem II.[60]
Trace gas analysis
Several techniques use ions created in a dedicated ion source injected into a flow tube or a drift tube: selected ion flow tube (SIFT-MS), and proton transfer reaction (PTR-MS), are variants of chemical ionization dedicated for trace gas analysis of air, breath or liquid headspace using well defined reaction time allowing calculations of analyte concentrations from the known reaction kinetics without the need for internal standard or calibration.
Another technique with applications in trace gas analysis field is secondary electrospray ionization (SESI-MS), which is a variant of electrospray ionization. SESI consist of an electrospray plume of pure acidified solvent that interacts with neutral vapors. Vapor molecules get ionized at atmospheric pressure when charge is transferred from the ions formed in the electrospray to the molecules. One advantage of this approach is that it is compatible with most ESI-MS systems.[61][62]
Residual gas analysis

A residual gas analyzer (RGA) is a small and usually rugged mass spectrometer, typically designed for process control and contamination monitoring in vacuum systems. When constructed as a quadrupole mass analyzer, there exist two implementations, utilizing either an open ion source (OIS) or a closed ion source (CIS). RGAs may be found in high vacuum applications such as research chambers, surface science setups, accelerators, scanning microscopes, etc. RGAs are used in most cases to monitor the quality of the vacuum and easily detect minute traces of impurities in the low-pressure gas environment. These impurities can be measured down to Torr levels, possessing sub-ppm detectability in the absence of background interferences.
RGAs would also be used as sensitive in-situ leak detectors commonly using helium, isopropyl alcohol or other tracer molecules. With vacuum systems pumped down to lower than Torr—checking of the integrity of the vacuum seals and the quality of the vacuum—air leaks, virtual leaks and other contaminants at low levels may be detected before a process is initiated.
Atom probe
An atom probe is an instrument that combines time-of-flight mass spectrometry and field-evaporation microscopy to map the location of individual atoms.
Pharmacokinetics
Pharmacokinetics is often studied using mass spectrometry because of the complex nature of the matrix (often blood or urine) and the need for high sensitivity to observe low dose and long time point data. The most common instrumentation used in this application is LC-MS with a triple quadrupole mass spectrometer. Tandem mass spectrometry is usually employed for added specificity. Standard curves and internal standards are used for quantitation of usually a single pharmaceutical in the samples. The samples represent different time points as a pharmaceutical is administered and then metabolized or cleared from the body. Blank or t=0 samples taken before administration are important in determining background and ensuring data integrity with such complex sample matrices. Much attention is paid to the linearity of the standard curve; however it is not uncommon to use curve fitting with more complex functions such as quadratics since the response of most mass spectrometers is less than linear across large concentration ranges.[63][64][65]
There is currently considerable interest in the use of very high sensitivity mass spectrometry for microdosing studies, which are seen as a promising alternative to animal experimentation.
Recent studies show that secondary electrospray ionization (SESI) is a powerful technique to monitor drug kinetics via breath analysis.[66][67] Because breath is naturally produced, several datapoints can be readily collected. This allows for the number of collected data-points to be greatly increased.[68] In animal studies, this approach SESI can reduce animal sacrifice.[67] In humans, SESI-MS non-invasive analysis of breath can help study the kinetics of drugs at a personalized level.[66][69][70]
Protein characterization
Mass spectrometry is an important method for the characterization and sequencing of proteins. The two primary methods for ionization of whole proteins are electrospray ionization (ESI) and matrix-assisted laser desorption/ionization (MALDI). In keeping with the performance and mass range of available mass spectrometers, two approaches are used for characterizing proteins. In the first, intact proteins are ionized by either of the two techniques described above, and then introduced to a mass analyzer. This approach is referred to as “top-down” strategy of protein analysis. The top-down approach however is largely limited to low-throughput single-protein studies. In the second, proteins are enzymatically digested into smaller peptides using proteases such as trypsin or pepsin, either in solution or in gel after electrophoretic separation. Other proteolytic agents are also used. The collection of peptide products are often separated by chromatography prior to introduction to the mass analyzer. When the characteristic pattern of peptides is used for the identification of the protein the method is called peptide mass fingerprinting (PMF), if the identification is performed using the sequence data determined in tandem MS analysis it is called de novo peptide sequencing. These procedures of protein analysis are also referred to as the “bottom-up” approach, and have also been used to analyse the distribution and position of post-translational modifications such as phosphorylation on proteins.[71] A third approach is also beginning to be used, this intermediate “middle-down” approach involves analyzing proteolytic peptides that are larger than the typical tryptic peptide.[72]
Space exploration

As a standard method for analysis, mass spectrometers have reached other planets and moons. Two were taken to Mars by the Viking program. In early 2005 the Cassini–Huygens mission delivered a specialized GC-MS instrument aboard the Huygens probe through the atmosphere of Titan, the largest moon of the planet Saturn. This instrument analyzed atmospheric samples along its descent trajectory and was able to vaporize and analyze samples of Titan’s frozen, hydrocarbon covered surface once the probe had landed. These measurements compare the abundance of isotope(s) of each particle comparatively to earth’s natural abundance.[73] Also on board the Cassini–Huygens spacecraft was an ion and neutral mass spectrometer which had been taking measurements of Titan’s atmospheric composition as well as the composition of Enceladus‘ plumes. A Thermal and Evolved Gas Analyzer mass spectrometer was carried by the Mars Phoenix Lander launched in 2007.[74]
Mass spectrometers are also widely used in space missions to measure the composition of plasmas. For example, the Cassini spacecraft carried the Cassini Plasma Spectrometer (CAPS),[75] which measured the mass of ions in Saturn’s magnetosphere.
Respired gas monitor
Mass spectrometers were used in hospitals for respiratory gas analysis beginning around 1975 through the end of the century. Some are probably still in use but none are currently being manufactured.[76]
Found mostly in the operating room, they were a part of a complex system, in which respired gas samples from patients undergoing anesthesia were drawn into the instrument through a valve mechanism designed to sequentially connect up to 32 rooms to the mass spectrometer. A computer directed all operations of the system. The data collected from the mass spectrometer was delivered to the individual rooms for the anesthesiologist to use.
The uniqueness of this magnetic sector mass spectrometer may have been the fact that a plane of detectors, each purposely positioned to collect all of the ion species expected to be in the samples, allowed the instrument to simultaneously report all of the gases respired by the patient. Although the mass range was limited to slightly over 120 u, fragmentation of some of the heavier molecules negated the need for a higher detection limit.[77]
Preparative mass spectrometry
The primary function of mass spectrometry is as a tool for chemical analyses based on detection and quantification of ions according to their mass-to-charge ratio. However, mass spectrometry also shows promise for material synthesis.[59] Ion soft landing is characterized by deposition of intact species on surfaces at low kinetic energies which precludes the fragmentation of the incident species.[78] The soft landing technique was first reported in 1977 for the reaction of low energy sulfur containing ions on a lead surface.[79]
Charge detection mass spectrometry
Most mass spectrometers measure the mass-to-charge ratio; the actual mass can be found only if the charge is known. For smaller molecules the charge can be determined from the spacing of isotope peaks, but for very large biomolecules and particles (in the megadalton range) resolution may not be adequate to separate isotope peaks, and thus the mass cannot be determined. In charge detection mass spectrometry (CDMS), the charge of an individual ion/particle is measured directly (alongside its mass-to-charge ratio) and therefore the true mass is known. It is a single-particle technique, but to produce more precise and accurate results, the data from many individually-measured ions can be combined.[80][81]
See also
- Bioelectrospray
- Dumas method of molecular weight determination
- Evolved gas analysis
- Helium mass spectrometer
- Isotope dilution
- MassBank (database), a Japanese spectral database
- Mass spectrometry imaging
- Mass spectrometry software
- MasSpec Pen
- Micro-arrays for mass spectrometry
- Nanoscale secondary ion mass spectrometry
- Reflectron
References
- ^ a b c Sparkman, O. David (2000). Mass spectrometry desk reference. Pittsburgh: Global View Pub. ISBN 978-0-9660813-2-9.
- ^ “Definition of spectrograph[permanent dead link].” Merriam Webster. Accessed 13 June 2008.
- ^ Downard K (2004). Mass Spectrometry – A Foundation Course. Royal Society of Chemistry. doi:10.1039/9781847551306. ISBN 978-0-85404-609-6.
- ^ Squires G (1998). “Francis Aston and the mass spectrograph”. Dalton Transactions (23): 3893–3900. doi:10.1039/a804629h.
- ^ Downard KM (2007). “Historical account: Francis William Aston: the man behind the mass spectrograph”. European Journal of Mass Spectrometry. 13 (3): 177–90. doi:10.1255/ejms.878. PMID 17881785. S2CID 25747367.
- ^ Thomson JJ (1913). Rays Of Positive Electricity and Their Application to Chemical Analysis. London: Longman’s Green and Company.
- ^ Siri W (1947). “Mass spectroscope for analysis in the low-mass range”. Review of Scientific Instruments. 18 (8): 540–545. Bibcode:1947RScI…18..540S. doi:10.1063/1.1740998.
- ^ Price P (August 1991). “Standard definitions of terms relating to mass spectrometry : A report from the committee on measurements and standards of the American society for mass spectrometry”. Journal of the American Society for Mass Spectrometry. 2 (4): 336–48. Bibcode:1991JASMS…2..336P. doi:10.1016/1044-0305(91)80025-3. PMID 24242353. S2CID 20236081.
- ^ Parkins WE (2005). “The uranium bomb, the calutron, and the space-charge problem”. Physics Today. 58 (5): 45–51. Bibcode:2005PhT….58e..45P. CiteSeerX 10.1.1.579.4119. doi:10.1063/1.1995747. ISSN 0031-9228.
- ^ “The Nobel Prize in Chemistry 2002: Information for the Public”. The Nobel Foundation. 9 October 2002. Retrieved 2007-08-29.
- ^ a b c d e f g h i j k l m n o p q r s t u v w x y z Pavia DL, Lampman GM, Kriz GS, Vyvyan JR (2015). Introduction to spectroscopy (5th ed.). Stamford, CT: Cengage Learning. ISBN 978-1-285-46012-3.
- ^ Fenn JB, Mann M, Meng CK, Wong SF, Whitehouse CM (October 1989). “Electrospray ionization for mass spectrometry of large biomolecules”. Science. 246 (4926): 64–71. Bibcode:1989Sci…246…64F. CiteSeerX 10.1.1.522.9458. doi:10.1126/science.2675315. PMID 2675315.
- ^ Tanaka K, Waki H, Ido Y, Akita S, Yoshida Y, Yoshida T (1988). “Protein and Polymer Analyses up to m/z 100 000 by Laser Ionization Time-of flight Mass Spectrometry”. Rapid Commun Mass Spectrom. 2 (20): 151–3. Bibcode:1988RCMS….2..151T. doi:10.1002/rcm.1290020802.
- ^ Karas M, Bachman D, Bahr U, Hillenkamp F (1987). “Matrix-Assisted Ultraviolet Laser Desorption of Non-Volatile Compounds”. Int J Mass Spectrom Ion Proc. 78: 53–68. Bibcode:1987IJMSI..78…53K. doi:10.1016/0168-1176(87)87041-6.
- ^ Northen TR, Yanes O, Northen MT, Marrinucci D, Uritboonthai W, Apon J, Golledge SL, Nordström A, Siuzdak G (October 2007). “Clathrate nanostructures for mass spectrometry”. Nature. 449 (7165): 1033–1036. Bibcode:2007Natur.449.1033N. doi:10.1038/nature06195. ISSN 1476-4687. PMID 17960240. S2CID 4404703.
- ^ Osborn DL, Zou P, Johnsen H, Hayden CC, Taatjes CA, Knyazev VD, North SW, Peterka DS, Ahmed M, Leone SR (October 2008). “The multiplexed chemical kinetic photoionization mass spectrometer: a new approach to isomer-resolved chemical kinetics”. The Review of Scientific Instruments (Submitted manuscript). 79 (10) 104103: 104103–104103–10. Bibcode:2008RScI…79j4103O. doi:10.1063/1.3000004. OSTI 953768. PMID 19044733. S2CID 25452748.
- ^ Bruins, A. P. (1991). “Mass spectrometry with ion sources operating at atmospheric pressure”. Mass Spectrometry Reviews. 10 (1): 53–77. Bibcode:1991MSRv…10…53B. doi:10.1002/mas.1280100104.
- ^ Cottrell JS, Greathead RJ (1986). “Extending the Mass Range of a Sector Mass Spectrometer”. Mass Spectrometry Reviews. 5 (3): 215–247. Bibcode:1986MSRv….5..215C. doi:10.1002/mas.1280050302.
- ^ “Gas Chromatography-Mass Spectrometry”. American Chemical Society. National Historic Chemical Landmarks. 2019-06-08.
- ^ Guilhaus M (1998). “Principles and Instrumentation in Time-of-flight Mass Spectrometry” (PDF). Journal of Mass Spectrometry. 30 (11): 1519–1532. doi:10.1002/jms.1190301102. S2CID 9444467. Archived from the original (PDF) on 2018-02-06 – via Google Scholar.
- ^ Wollnik, H. (1993). “Time-of-flight mass analyzers”. Mass Spectrometry Reviews. 12 (2): 89–114. Bibcode:1993MSRv…12…89W. doi:10.1002/mas.1280120202.
- ^ Syed SU, Maher S, Taylor S (December 2013). “Quadrupole mass filter operation under the influence of magnetic field”. Journal of Mass Spectrometry. 48 (12): 1325–39. Bibcode:2013JMSp…48.1325S. doi:10.1002/jms.3293. PMID 24338888.
- ^ Maher S, Syed SU, Hughes DM, Gibson JR, Taylor S (August 2013). “Mapping the stability diagram of a quadrupole mass spectrometer with a static transverse magnetic field applied”. Journal of the American Society for Mass Spectrometry. 24 (8): 1307–14. Bibcode:2013JASMS..24.1307M. doi:10.1007/s13361-013-0654-5. PMID 23720050. S2CID 45734248.
- ^ Paul W, Steinwedel H (1953). “Ein neues Massenspektrometer ohne Magnetfeld”. Zeitschrift für Naturforschung A. 8 (7): 448–450. Bibcode:1953ZNatA…8..448P. doi:10.1515/zna-1953-0710. S2CID 96549388.
- ^ March RE (2000). “Quadrupole ion trap mass spectrometry: a view at the turn of the century”. International Journal of Mass Spectrometry. 200 (1–3): 285–312. Bibcode:2000IJMSp.200..285M. doi:10.1016/S1387-3806(00)00345-6.
- ^ Schwartz JC, Senko MW, Syka JE (June 2002). “A two-dimensional quadrupole ion trap mass spectrometer”. Journal of the American Society for Mass Spectrometry. 13 (6): 659–69. Bibcode:2002JASMS..13..659S. doi:10.1016/S1044-0305(02)00384-7. PMID 12056566. S2CID 26965687.
- ^ Lammert SA, Rockwood AA, Wang M, Lee ML, Lee ED, Tolley SE, Oliphant JR, Jones JL, Waite RW (July 2006). “Miniature toroidal radio frequency ion trap mass analyzer”. Journal of the American Society for Mass Spectrometry. 17 (7): 916–922. Bibcode:2006JASMS..17..916L. doi:10.1016/j.jasms.2006.02.009. PMID 16697659. S2CID 45444397.
- ^ Snyder DT, Pulliam CJ, Ouyang Z, Cooks RG (January 2016). “Miniature and Fieldable Mass Spectrometers: Recent Advances”. Analytical Chemistry. 88 (1): 2–29. Bibcode:2016AnaCh..88….2S. doi:10.1021/acs.analchem.5b03070. PMC 5364034. PMID 26422665.
- ^ Comisarow MB, Marshall AG (1974). “Fourier-transform ion cyclotron resonance spectroscopy”. Chemical Physics Letters. 25 (2): 282–283. Bibcode:1974CPL….25..282C. doi:10.1016/0009-2614(74)89137-2.
- ^ Marshall AG, Hendrickson CL, Jackson GS (1998). “Fourier-transform ion cyclotron resonance mass spectrometry: a primer”. Mass Spectrometry Reviews. 17 (1): 1–35. Bibcode:1998MSRv…17….1M. doi:10.1002/(SICI)1098-2787(1998)17:1<1::AID-MAS1>3.0.CO;2-K. PMID 9768511.
- ^ Hu Q, Noll RJ, Li H, Makarov A, Hardman M, Graham Cooks R (April 2005). “The Orbitrap: a new mass spectrometer”. Journal of Mass Spectrometry. 40 (4): 430–43. Bibcode:2005JMSp…40..430H. doi:10.1002/jms.856. PMID 15838939.
- ^ a b c Smith DF, Podgorski DC, Rodgers RP, Blakney GT, Hendrickson CL (2018-02-06). “21 Tesla FT-ICR Mass Spectrometer for Ultrahigh-Resolution Analysis of Complex Organic Mixtures”. Analytical Chemistry. 90 (3): 2041–2047. doi:10.1021/acs.analchem.7b04159. ISSN 0003-2700.
- ^ Xian F, Hendrickson CL, Marshall AG (2012-01-17). “High Resolution Mass Spectrometry”. Analytical Chemistry. 84 (2): 708–719. doi:10.1021/ac203191t. ISSN 0003-2700.
- ^ Marshall AG, Guan S (1996). “Advantages of High Magnetic Field for Fourier Transform Ion Cyclotron Resonance Mass Spectrometry”. Rapid Communications in Mass Spectrometry. 10 (14): 1819–1823. doi:10.1002/(SICI)1097-0231(199611)10:14<1819::AID-RCM686>3.0.CO;2-Z. ISSN 1097-0231.
- ^ a b Shaw JB, Lin TY, Leach FE, Tolmachev AV, Tolić N, Robinson EW, Koppenaal DW, Paša-Tolić L (2016-12-01). “21 Tesla Fourier Transform Ion Cyclotron Resonance Mass Spectrometer Greatly Expands Mass Spectrometry Toolbox”. Journal of the American Society for Mass Spectrometry. 27 (12): 1929–1936. doi:10.1007/s13361-016-1507-9.
- ^ Bogdanov B, Smith RD (March 2005). “Proteomics by FTICR mass spectrometry: Top down and bottom up”. Mass Spectrometry Reviews. 24 (2): 168–200. doi:10.1002/mas.20015. ISSN 0277-7037.
- ^ Dubois F, Knochenmuss R, Zenobi R, Brunelle A, Deprun C, Le Beyec Y (1999). “A comparison between ion-to-photon and microchannel plate detectors”. Rapid Communications in Mass Spectrometry. 13 (9): 786–791. Bibcode:1999RCMS…13..786D. doi:10.1002/(SICI)1097-0231(19990515)13:9<786::AID-RCM566>3.0.CO;2-3.
- ^ Park MA, Callahan JH, Vertes A (1994). “An inductive detector for time-of-flight mass spectrometry”. Rapid Communications in Mass Spectrometry. 8 (4): 317–322. Bibcode:1994RCMS….8..317P. doi:10.1002/rcm.1290080407.
- ^ Boyd, Robert K. (1994). “Linked-scan techniques for MS/MS using tandem-in-space instruments”. Mass Spectrometry Reviews. 13 (5–6): 359–410. Bibcode:1994MSRv…13..359B. doi:10.1002/mas.1280130502.
- ^ Busch KL. “A Glossary for Mass Spectrometry” (PDF). Waters.com. Waters. Retrieved 12 February 2024.
- ^ a b c Xue J, Guijas C, Benton HP, Warth B, Siuzdak G (October 2020). “METLIN MS2 molecular standards database: a broad chemical and biological resource”. Nature Methods. 17 (10): 953–954. Bibcode:2020NatCB..17..953X. doi:10.1038/s41592-020-0942-5. PMC 8802982. PMID 32839599. S2CID 221285246.
- ^ Smith CA, O’Maille G, Want EJ, Qin C, Trauger SA, Brandon TR, Custodio DE, Abagyan R, Siuzdak G (December 2005). “METLIN: a metabolite mass spectral database”. Therapeutic Drug Monitoring. 27 (6): 747–51. doi:10.1097/01.ftd.0000179845.53213.39. PMID 16404815. S2CID 14774455.
- ^ Guijas C, Montenegro-Burke JR, Domingo-Almenara X, Palermo A, Warth B, Hermann G, Koellensperger G, Huan T, Uritboonthai W, Aisporna AE, Wolan DW, Spilker ME, Benton HP, Siuzdak G (March 2018). “METLIN: A Technology Platform for Identifying Knowns and Unknowns”. Analytical Chemistry. 90 (5): 3156–3164. Bibcode:2018AnaCh..90.3156G. doi:10.1021/acs.analchem.7b04424. PMC 5933435. PMID 29381867.
- ^ a b Heim W, Aisporna A, Hoang L, Benton HP, Siuzdak G (2023), Ivanisevic J, Giera M (eds.), “METLIN Tandem Mass Spectrometry and Neutral Loss Databases for the Identification of Microbial Natural Products and Other Chemical Entities”, A Practical Guide to Metabolomics Applications in Health and Disease: From Samples to Insights into Metabolism, Cham: Springer International Publishing, pp. 105–124, doi:10.1007/978-3-031-44256-8_5, ISBN 978-3-031-44256-8, retrieved 2024-03-27
{{citation}}: CS1 maint: work parameter with ISBN (link) - ^ Aisporna A, Webb B, Uritboonthai W, Hoang L, Billings EM, Hoang C, Plumb R, Siuzdak G (2025-10-28). “Adaptive Modeling of Tandem Mass Spectrometry Data: Creation of the METLIN 960K MRM Database”. Analytical Chemistry. 97 (44): 24586–24595. Bibcode:2025AnaCh..9724586A. doi:10.1021/acs.analchem.5c04639. ISSN 0003-2700. PMC 12613142. PMID 41150651.
- ^ Hoang C, Uritboonthai W, Hoang L, Billings EM, Aisporna A, Nia FA, Derks RJ, Williamson JR, Giera M, Siuzdak G (2024-03-26). “Tandem Mass Spectrometry across Platforms”. Analytical Chemistry. 96 (14): 5478–5488. Bibcode:2024AnaCh..96.5478H. doi:10.1021/acs.analchem.3c05576. ISSN 0003-2700. PMC 11007677. PMID 38529642.
- ^ Benton HP, Wong DM, Trauger SA, Siuzdak G (August 2008). “XCMS2: processing tandem mass spectrometry data for metabolite identification and structural characterization”. Analytical Chemistry. 80 (16): 6382–9. Bibcode:2008AnaCh..80.6382B. doi:10.1021/ac800795f. PMC 2728033. PMID 18627180.
- ^ Eiceman, G.A. (2000). Gas Chromatography. In R.A. Meyers (Ed.), Encyclopedia of Analytical Chemistry: Applications, Theory, and Instrumentation, pp. 10627. Chichester: Wiley. ISBN 0-471-97670-9
- ^ Fang M, Ivanisevic J, Benton HP, Johnson CH, Patti GJ, Hoang LT, Uritboonthai W, Kurczy ME, Siuzdak G (November 2015). “Thermal Degradation of Small Molecules: A Global Metabolomic Investigation”. Analytical Chemistry. 87 (21): 10935–41. Bibcode:2015AnaCh..8710935F. doi:10.1021/acs.analchem.5b03003. PMC 4633772. PMID 26434689.
- ^ Loo JA, Udseth HR, Smith RD (June 1989). “Peptide and protein analysis by electrospray ionization-mass spectrometry and capillary electrophoresis-mass spectrometry”. Analytical Biochemistry. 179 (2): 404–12. doi:10.1016/0003-2697(89)90153-X. PMID 2774189.
- ^ Maxwell EJ, Chen DD (October 2008). “Twenty years of interface development for capillary electrophoresis-electrospray ionization-mass spectrometry”. Analytica Chimica Acta. 627 (1): 25–33. Bibcode:2008AcAC..627…25M. doi:10.1016/j.aca.2008.06.034. PMID 18790125.
- ^ Verbeck GF, Ruotolo BT, Sawyer HA, Gillig KJ, Russell DH (June 2002). “A fundamental introduction to ion mobility mass spectrometry applied to the analysis of biomolecules”. Journal of Biomolecular Techniques. 13 (2): 56–61. PMC 2279851. PMID 19498967.
- ^ Matz LM, Asbury GR, Hill HH (2002). “Two-dimensional separations with electrospray ionization ambient pressure high-resolution ion mobility spectrometry/quadrupole mass spectrometry”. Rapid Communications in Mass Spectrometry. 16 (7): 670–5. Bibcode:2002RCMS…16..670M. doi:10.1002/rcm.623. PMID 11921245.
- ^ Sowell RA, Koeniger SL, Valentine SJ, Moon MH, Clemmer DE (September 2004). “Nanoflow LC/IMS-MS and LC/IMS-CID/MS of protein mixtures”. Journal of the American Society for Mass Spectrometry. 15 (9): 1341–53. Bibcode:2004JASMS..15.1341S. doi:10.1016/j.jasms.2004.06.014. PMID 15337515. S2CID 11531292.
- ^ Tureček F, McLafferty FW (1993). Interpretation of mass spectra. Sausalito: University Science Books. ISBN 978-0-935702-25-5.
- ^ Mistrik R. “A New Concept for the Interpretation of Mass Spectra Based on a Combination of a Fragmentation Mechanism Database and a Computer Expert System”. Highchem.com. Archived from the original on 11 January 2012.
- ^ Sheldon MT, Mistrik R, Croley TR (March 2009). “Determination of ion structures in structurally related compounds using precursor ion fingerprinting”. Journal of the American Society for Mass Spectrometry. 20 (3): 370–6. Bibcode:2009JASMS..20..370S. doi:10.1016/j.jasms.2008.10.017. PMID 19041260.
- ^ Zhou B, Xiao JF, Tuli L, Ressom HW (2012). “LC-MS-based metabolomics”. Mol. BioSyst. 8 (2): 470–481. doi:10.1039/c1mb05350g. ISSN 1742-206X. PMC 3699692. PMID 22041788.
- ^ a b Maher S, Jjunju FP, Taylor S (2015). “100 years of mass spectrometry: Perspectives and future trends”. Rev. Mod. Phys. 87 (1): 113–135. Bibcode:2015RvMP…87..113M. doi:10.1103/RevModPhys.87.113.
- ^ Shevela D, Messinger J (November 2013). “Studying the oxidation of water to molecular oxygen in photosynthetic and artificial systems by time-resolved membrane-inlet mass spectrometry”. Frontiers in Plant Science. 4: 473. Bibcode:2013FrPS….4..473S. doi:10.3389/fpls.2013.00473. PMC 3840314. PMID 24324477.
- ^ Li X, Huang L, Zhu H, Zhou Z (February 2017). “Direct human breath analysis by secondary nano-electrospray ionization ultrahigh-resolution mass spectrometry: Importance of high mass resolution and mass accuracy”. Rapid Communications in Mass Spectrometry. 31 (3): 301–308. Bibcode:2017RCMS…31..301L. doi:10.1002/rcm.7794. PMID 27859758.
- ^ Barrios-Collado C, Vidal-de-Miguel G, Martinez-Lozano Sinues P (February 2016). “Numerical modeling and experimental validation of a universal secondary electrospray ionization source for mass spectrometric gas analysis in real-time”. Sensors and Actuators B: Chemical. 223: 217–225. Bibcode:2016SeAcB.223..217B. doi:10.1016/j.snb.2015.09.073. hdl:20.500.11850/105470.
- ^ Hsieh Y, Korfmacher WA (June 2006). “Increasing speed and throughput when using HPLC-MS/MS systems for drug metabolism and pharmacokinetic screening”. Current Drug Metabolism. 7 (5): 479–89. doi:10.2174/138920006777697963. PMID 16787157. S2CID 13612670.
- ^ Covey TR, Lee ED, Henion JD (October 1986). “High-speed liquid chromatography/tandem mass spectrometry for the determination of drugs in biological samples”. Analytical Chemistry. 58 (12): 2453–60. Bibcode:1986AnaCh..58.2453C. doi:10.1021/ac00125a022. PMID 3789400.
- ^ Covey TR, Crowther JB, Dewey EA, Henion JD (February 1985). “Thermospray liquid chromatography/mass spectrometry determination of drugs and their metabolites in biological fluids”. Analytical Chemistry. 57 (2): 474–81. Bibcode:1985AnaCh..57..474C. doi:10.1021/ac50001a036. PMID 3977076.
- ^ a b Gamez G, Zhu L, Disko A, Chen H, Azov V, Chingin K, Krämer G, Zenobi R (May 2011). “Real-time, in vivo monitoring and pharmacokinetics of valproic acid via a novel biomarker in exhaled breath”. Chemical Communications. 47 (17): 4884–6. doi:10.1039/c1cc10343a. PMID 21373707.
- ^ a b Li X, Martinez-Lozano Sinues P, Dallmann R, Bregy L, Hollmén M, Proulx S, Brown SA, Detmar M, Kohler M, Zenobi R (June 2015). “Drug Pharmacokinetics Determined by Real-Time Analysis of Mouse Breath”. Angewandte Chemie. 54 (27): 7815–8. Bibcode:2015ACIE…54.7815L. doi:10.1002/anie.201503312. hdl:20.500.11850/102558. PMID 26015026.
- ^ Gaugg MT, Engler A, Nussbaumer-Ochsner Y, Bregy L, Stöberl AS, Gaisl T, Bruderer T, Zenobi R, Kohler M, Martinez-Lozano Sinues P (September 2017). “Metabolic effects of inhaled salbutamol determined by exhaled breath analysis”. Journal of Breath Research. 11 (4): 046004. Bibcode:2017JBR….11d6004G. doi:10.1088/1752-7163/aa7caa. hdl:20.500.11850/220016. PMID 28901297.
- ^ Martinez-Lozano Sinues P, Kohler M, Brown SA, Zenobi R, Dallmann R (February 2017). “Gauging circadian variation in ketamine metabolism by real-time breath analysis”. Chemical Communications. 53 (14): 2264–2267. doi:10.1039/C6CC09061C. PMID 28150005.
- ^ Tejero Rioseras A, Singh KD, Nowak N, Gaugg MT, Bruderer T, Zenobi R, Sinues PM (June 2018). “Real-Time Monitoring of Tricarboxylic Acid Metabolites in Exhaled Breath”. Analytical Chemistry. 90 (11): 6453–6460. Bibcode:2018AnaCh..90.6453T. doi:10.1021/acs.analchem.7b04600. PMID 29767961.
- ^ Ferries S, Perkins S, Brownridge PJ, Campbell A, Eyers PA, Jones AR, Eyers CE (September 2017). “Evaluation of Parameters for Confident Phosphorylation Site Localization Using an Orbitrap Fusion Tribrid Mass Spectrometer”. Journal of Proteome Research. 16 (9): 3448–3459. doi:10.1021/acs.jproteome.7b00337. PMID 28741359.
- ^ Chait BT (2011). “Mass spectrometry in the postgenomic era”. Annual Review of Biochemistry. 80: 239–46. doi:10.1146/annurev-biochem-110810-095744. PMID 21675917. S2CID 2676180. – via Annual Reviews (subscription required)
- ^ Petrie S, Bohme DK (2007). “Ions in space”. Mass Spectrometry Reviews. 26 (2): 258–80. Bibcode:2007MSRv…26..258P. doi:10.1002/mas.20114. PMID 17111346.
- ^ Hoffman JH, Chaney RC, Hammack H (October 2008). “Phoenix Mars Mission–the thermal evolved gas analyzer”. Journal of the American Society for Mass Spectrometry. 19 (10): 1377–83. Bibcode:2008JASMS..19.1377H. doi:10.1016/j.jasms.2008.07.015. PMID 18715800.
- ^ “Cassini Plasma Spectrometer”. Southwest Research Institute. Archived from the original on 2018-10-08. Retrieved 2008-01-04.
- ^ Riker JB, Haberman B (1976). “Expired gas monitoring by mass spectrometry in a respiratory intensive care unit”. Critical Care Medicine. 4 (5): 223–9. doi:10.1097/00003246-197609000-00002. PMID 975846. S2CID 6334599.
- ^ Gothard JW, Busst CM, Branthwaite MA, Davies NJ, Denison DM (September 1980). “Applications of respiratory mass spectrometry to intensive care”. Anaesthesia. 35 (9): 890–5. doi:10.1111/j.1365-2044.1980.tb03950.x. PMID 6778243. S2CID 41696370.
- ^ Verbeck G, Hoffmann W, Walton B (October 2012). “Soft-landing preparative mass spectrometry”. The Analyst. 137 (19): 4393–407. Bibcode:2012Ana…137.4393V. doi:10.1039/C2AN35550G. PMID 22900257.
- ^ Franchetti V, Solka BH, Baitinger WE, Amy JW, Cooks RG (1977). “Soft landing of ions as a means of surface modification”. Mass Spectrom. Ion Phys. 23 (1): 29–35. Bibcode:1977IJMSI..23…29F. doi:10.1016/0020-7381(77)80004-1.
- ^ Jarrold MF (2022). “Applications of Charge Detection Mass Spectrometry in Molecular Biology and Biotechnology”. Chemical Reviews. 122 (8): 7415–7441. doi:10.1021/acs.chemrev.1c00377. PMC 10842748. PMID 34637283.
- ^ Rathore AS (2022). “Charge Detection Mass Spectrometry: What’s the “Big” Deal?”. LCGC North America. 40 (3): 125–127. doi:10.56530/lcgc.na.gp8165g1. Retrieved 19 August 2025.
Bibliography
- Tureček F, McLafferty FW (1993). Interpretation of mass spectra. Sausalito, Calif: University Science Books. ISBN 978-0-935702-25-5.
- de Hoffman E, Stroobant V (2001). Mass Spectrometry: Principles and Applications (2nd ed.). John Wiley and Sons. ISBN 978-0-471-48566-7.
- Downard K (2004). Mass Spectrometry – A Foundation Course. Cambridge UK: Royal Society of Chemistry. ISBN 978-0-85404-609-6.
- Siuzdak G (1996). Mass spectrometry for biotechnology. Boston: Academic Press. ISBN 978-0-12-647471-8.
- Dass C (2001). Principles and practice of biological mass spectrometry. New York: John Wiley. ISBN 978-0-471-33053-0.
- Muzikar P, et al. (2003). “Accelerator Mass Spectrometry in Geologic Research”. Geological Society of America Bulletin. 115: 643–654. Bibcode:2003GSAB..115..643M. doi:10.1130/0016-7606(2003)115<0643:AMSIGR>2.0.CO;2. ISSN 0016-7606. S2CID 55076131.
- Maher S, Jjunju FP, Taylor S (2015). “100 years of mass spectrometry: Perspectives and future trends”. Rev. Mod. Phys. 87 (1): 113–135. Bibcode:2015RvMP…87..113M. doi:10.1103/RevModPhys.87.113.
- Sobott F (2014). Biological Mass Spectrometry. Boca Raton: Crc Pr I Llc. ISBN 978-1-4398-9527-6.
- Sparkman OD (2006). Mass Spectrometry Desk Reference. Pittsburgh: Global View Pub. ISBN 978-0-9660813-9-8.
- Watson JT, Sparkman OD (2007). Introduction to Mass Spectrometry: Instrumentatio, Applications, and Strategies for Data Interpretation (4th ed.). Chichester: Jonh Wiley & Sons. ISBN 978-0-470-51634-8.
- Tuniz C (1998). Accelerator mass spectrometry: ultrasensitive analysis for global science. Boca Raton: CRC Press. ISBN 978-0-8493-4538-8.
- Kandiah M, Urban PL (June 2013). “Advances in ultrasensitive mass spectrometry of organic molecules”. Chemical Society Reviews. 42 (12): 5299–322. doi:10.1039/c3cs35389c. PMID 23471277.
- Calmes J (2011). Mass spec : the biography of a scientific instrument (MS). Massachusetts Instite of Technology. hdl:1721.1/68473.
External links
- Interactive tutorial on mass spectra National High Magnetic Field Laboratory
- Mass spectrometer simulation An interactive application simulating the console of a mass spectrometer
- Realtime Mass Spectra simulation Tool to simulate mass spectra in the browser